Basic Information

Recording

Printer Paper

Revision D
Specifications: Basic Information
Waveforms and alphanumeric characters are recoreded directly with rectangular
coordinates by means of thermal-array printhead printing on thermosensitive paper.
Three or six recording channels, or 12 in 12 Lead Mode, overlapping
Baseline pitch 3 channels: 62 mm (arrhythmia) 
6 channels: 31 mm (6 Lead)
12 channels: 16 mm (12 Lead.)
Writing width 200 mm maximum
Recorder settings, date, time and entered patient name are annotated in the
margin of the recording strip
With appropriate software, documentation of analysis results in the respective
operating mode
Resolution of the recording.
Vertical: 8 dots/mm
Horizontal: 25 mm at 25 mm/s
GE CONTRAST Z-fold pad, 150 pages per pad, equivalent to a chart length of
approximately 45 m.
Paper width: 8.5 inch
Sheet length: 11 inch
To prevent damage to the printhead use only the original GE CONTRAST paper or
the GE thermal paper with queue holes or marks.
MAC 1200 resting ECG analysis system
2012250-095
9-3