Safety Instructions

Revision D
Repair Instructions: Safety Instructions
When repairing the MAC 1200 resting ECG analysis system unit, the following
considerations apply:
Repairs may only be conducted by approved service personnel.
Before opening the MAC 1200 resting ECG analysis system, switch off
device and disconnect the mains plug.
Never connect the mains plug when the device is open.
Before replacing the primary fuses in the power input module, the device
should also be switched off and the mains plug disconnected.
For replacing components, only the original GE components, mentioned in the
spare parts list may be used.
When replacing electronic components, implement ESD protection.
Return replaced PCBs in ESD packaging only.
Defective NC batteries or empty batteries should be disposed of in accordance
with the applicable legal stipulations or returned to the factory.
Batteries returned to the factory should be labeled For disposal.
MAC 1200 resting ECG analysis system
2012250-095
4-3