System Test Function: Pharma
The MAC 1200 device variant is configured in the service mode and any MAC 1200
can be configured as a PHARMA device.The Pharma device variant is designed for
use by Clinical Research Organizations (CROs) in pharmaceutical trials. MAC 1200
units with the Pharma option are configured for specific pharmaceutical trials by
CROs and then sent out to trial sites to collect data.
The Pharma option of the MAC 1200 can be password-protected (HIGH
SECURITY). Password protection is enabled in the system setup.
NOTE
The Pharma and HIGH SECURITY options appear only if the device has been
configured as a Pharma device.
High Security provides the option of password-protecting use of the MAC 1200
unit. Any device configured with the Pharma option may enable the High Security
option.
When the High Security option is enabled, users must provide information for
the Technician Name field.
To bypass the logon password, press