To set the DICOM properties for your LOGIQworks system:
1.) Select Configuration > Generic... from the Data Selector menu.
2.) Select the DICOM tab from the Generic Properties dialog.
These steps describe how to enter the correct Application Entity Title and Port number.
AE Title: DICOM Application Entity title for the DICOM services provided by your system. The AE
Title entered here should be used by other systems to set up DICOM associations to your system.
The same AE Title is also used by your system when it sets up associations to other systems. The
AE Title is then referred to as the calling AE Title.
For LOGIQ 7 the AE Title has always to begin with "AE_" otherwise the images can not be transferred.
NOTE
Port No.: Socket port number for the DICOM services provided by this system. The Port Number
is used by other systems to set up DICOM associations to this system.(Default: 104)
Reboot entire system after any change to DICOM settings.
NOTE