GE LOGIQworks Basic Service Manual page 201

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
5184024-100, R
IRECTION
3-5-4-3
Logging (cont'd)
Figure 3-121 Connection Service Properties - Logging (Default Settings)
Logging Level: Set the level of logging messages. If set to ALL, all messages will be logged. If set
to NORMAL (the typical setting) only error messages will be logged.
Logging Type: Set the types of logging used by the RadWorks application. (Default: File)
-
-
-
Log File: Set the name of the log file. Each Service has its own default log filename. (Do not
change)
Truncation Size (kB): Set the truncation size of the logging file. The truncation size is the size the
log file will be shortened to when its maximum size has been exceeded. (Default: 800 kB)
Maximum Size (kB): Set the maximum size of the log file. (Default: 1024 kB)
5
EVISION
D:\RA600\LOG
File: If this check box is checked, RadWorks will write general information about its operation
to the RadWorks.log file, which you can view in the Log folder.
Pipe: Used in combination with the listening tool in the X:\AMI_61\bin folder. Among other
things, it enables you to observe DICOM transmissions in a pipe. Consult the technical
reference manual or contact technical support for additional information.
Console: If this check box is checked, RadWorks will display a window that continuously
shows the actions RadWorks is performing. This is useful when troubleshooting.
Chapter 3 Installation
LOGIQ
® B
S
WORKS
ASIC
ERVICE
M
ANUAL
3-135