GE LOGIQworks Basic Service Manual page 209

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
5184024-100, R
IRECTION
3-5-4-7
DICOM (cont'd)
Figure 3-125 Connection Service Properties - DICOM (Default settings)
The settings are divided into a section for DICOM Receive, followed by a section for DICOM send. The
DICOM Application Entity Title and Port Number can be set in the DICOM tab in Generic System
configuration (please refer to
Thread Priority: Priority of the DICOM service thread within the Connection Service. Changing this
setting can influence the system performance. (Default: normal)
DICOM Receive
-
-
5
EVISION
Section 3-5-2-6 "DICOM settings" on page
Strict DICOM Conformance: This flag can be used to force strict DICOM conformance when
receiving DICOM transfers from other systems. If images arrive that do not comply with strict
DICOM conformance, these images will not appear in the database. This is also logged in the
Receive Log. (Default: not checked)
Wait before Converting: This number can be used to set a time-out for DICOM conversion
when a DICOM transfer is done on an image-by-image basis, i.e. when for every image a new
association is started. If not zero, the number specifies (in milliseconds) the time-out between
images. (Default: 0)
Chapter 3 Installation
LOGIQ
® B
S
WORKS
ASIC
ERVICE
3-87).
M
ANUAL
3-143

Advertisement

Table of Contents
loading

Table of Contents