Seal On Pc Slide-In Module; Pc Diagnostic Tool; Starting The Diagnostic Tool - Sirona CEREC AC BLUECAM Operating Instructions Manual

Dental acquisition unit
Table of Contents

Advertisement

10 Appendix
10.3 Seal on PC slide-in module
66
10.3

Seal on PC slide-in module

CEREC AC graphics seal
NOTICE
Text seal
If the seal is broken, all warranty claims regarding the PC slide-module
automatically expire.
The PC slide-in module may be opened only by an authorized dental
technician. Only spare parts approved by us may be used in this module.
Following a repair, the seal supplied along with the spare parts must be
affixed at the specified location (A).
10.4

PC Diagnostic Tool

10.4.1

Starting the diagnostic tool

Test
Customer Diagnostics
Technician Diagnostics
Sirona Windows Diagnostics
Tip: If one of the following steps does not work, further steps can be found
in the Service Manual.
The PC is switched off.
1. Switch on the PC and wait until the above message appears.
2. Hold down the "F11" key on the keyboard until the boot menu
appears.
Sirona Dental Systems GmbH
Operating Instructions for the acquisition unit CEREC AC
Suitable for
Suitable for the user, in order to check
the PC components.
Suitable for the technician, in order to
check the PC components.
Windows Stress Test, in order to test
the PC at permanent load.
D3492.201.01.17.02 12.2015
62 31 174 D3492

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cerec ac

Table of Contents