System Settings

46
Operating Instructions Cubis
7
System Settings
7.1

Performing System Settings

Default settings can be adjusted for the device and the applications in order
to align with the ambient conditions and individual operating requirements.
The following settings are necessary to operate the device together with
connected components:
− Set up the communication of the connected devices
− Set up additional components
The following settings are recommended to set up the device:
− Set the menu language
− Select date | time format
− Setting the date, time, and time zone
− Assigning a password
− Set the behavior of the isoCAL function
Procedure
Open the main menu.
t
Press the [Setup] button.
t
To adjust settings: Open the desired submenu.
t
Select the desired configuration value (configuration values,
t
see Chapter "7.7 Parameter List", page 51).
Exit the menu.
t
7.2

Switching Off the isoCAL Function

If the isoCAL function is switched off for a conformity-assessed device:
The device can only be used for legal-for-trade applications in restricted
temperature ranges (see Chapter "15.3.2 Ambient Temperature for the
isoCAL Function", page 91). The isoCAL function cannot be switched
off for all model versions.
Procedure
In the "Settings / Device Settings / Safe Weighing / isoCAL Execution
t
Mode" menu, for the "isoCAL function" parameter, select the
"Off" configuration value.