High flow therapy
Changing from an NIV
mode to high flow
therapy
7-4

Connecting directly to a 22 mm circuit

Remove the DEP/FEP. Connect the high flow nasal cannula with 22 mm
connector directly to the circuit.
DEP/FEP
Figure 7-2: High flow nasal cannula, 22 mm connection
Follow these instructions to use the V60/V60 Plus Ventilator for high flow
therapy (HFT).
1. Select Standby. The Entering Standby window opens.
2. Remove the patient mask or ET interface to enter Standby.
3. Install a Philips-approved high flow nasal cannula (Figure 7-1 and
Figure 7-2 above) or a high flow tracheostomy interface on the patient
circuit.
4. Select
.
HFT
22 mm connector