Table 9-5: Vent Inoperative alarm messages: summary and troubleshooting
Message
Description
Vent Inoperative
Technical failure. The
1000
ventilator is in the ven-
tilator inoperative
3.3 V Supply
state.
Failed
Vent Inoperative
Technical failure. The
1001
ventilator is in the ven-
tilator inoperative
12 V Supply
state.
Failed
Vent Inoperative
Technical failure. The
1002
ventilator is in the ven-
tilator inoperative
Blower Tempera-
state.
ture Too High
Vent Inoperative
Technical failure of the
1003
CPU PCBA. The venti-
lator is in the ventilator
Internal Tempera-
inoperative state.
ture High
Vent Inoperative
Technical failure of the
1004
DAQ PCBA. The venti-
lator is in the ventilator
Internal Tempera-
inoperative state.
ture High
Vent Inoperative
Technical failure of the
1005
motor PCBA. The venti-
lator is in the ventilator
Internal Tempera-
inoperative state.
ture High
Vent Inoperative
Technical failure. The
1006
ventilator is in the ven-
tilator inoperative
Data Acquisition
state.
PCBA ADC Failed
Vent Inoperative
Technical failure. The
1007
ventilator is in the ven-
tilator inoperative
Machine and
state.
Proximal Pres-
sure Sensors
Failed
Vent Inoperative
Technical failure. The
1008
ventilator is in the ven-
tilator inoperative
Machine and
state.
Proximal Pres-
sure Sensors
Failed
Alarms, messages, and troubleshooting
Corrective Action
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
Priority
Manually
Autores
type (ID)
resettable
ettable
High
Yes
No
(2)
High
Yes
No
(3)
High
Yes
No
(4)
High
Yes
No
(5)
High
Yes
No
(6)
High
Yes
No
(7)
High
Yes
No
(8)
High
Yes
No
(9)
High
Yes
No
(10)
Silence
able
No
No
No
No
No
No
No
No
No
9-17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

V60 plus

Table of Contents