Alarms, messages, and troubleshooting
Table 9-4: Check Vent alarm messages: summary and troubleshooting (continued)
Message
Description
Check Vent: O
Pres-
Continues to ventilate
2
sure Sensor Calibra-
with air only
tion Data Error
Check Vent: O
Sup-
Continues to ventilate
2
ply Pressure Sensor
with air only
Range Error
Check Vent: OVP
Technical failure
Circuit Failed
Check Vent: Oxygen
Continues to ventilate
Device Failed
with air only
Check Vent: Primary
Technical failure
Alarm Failed
Check Vent: Pro-
Technical failure
gram CRC Test Failed
Check Vent: Proximal
Proximal pressure is
Pressure Sensor
not measured. Pres-
Auto-Zero Failed
sure-related alarms are
compromised.
Check Vent: Proximal
Proximal pressure is
Pressure Sensor Cali-
not measured. Pres-
bration Data Error
sure-related alarms are
compromised.
Check Vent: Proximal
Proximal pressure is
Pressure Sensor
not measured. Pres-
Range Error
sure-related alarms are
compromised.
Also displays when
HFT (high flow therapy)
is active and proximal
pressure is out of
range.
Check Vent: Ventila-
Technical failure
tor Restarted
9-16
Corrective Action
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
When using a high flow na-
sal cannula with FEP Con-
nect, the proximal pressure
line should be disconnected
from the ventilator port
during HFT. If this message
is seen when switching the
patient from ventilation to
HFT, check to make sure the
proximal pressure line is dis-
connected.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Priority
Manually
Autores
type (ID)
resettable
ettable
High
Yes
No
(15)
High
Yes
No
(17)
High
Yes
No
(38)
High
Yes
No
(16)
High
Yes
No
(4)
High
Yes
No
(1)
High
Yes
No
(10)
High
Yes
No
(8)
High
Yes
No
(12)
High
Yes
No
(3)
Silence
able
No
No
No
No
No
No
No
No
No
No

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

V60 plus

Table of Contents