Electrical
specifications
Accessory
requirements
11-8
Table 11-10: Electrical specifications
Parameter
AC voltage
AC frequency
AC power
Battery
NOTE:
Any part that is added to the patient circuit changes the compliance
and resistance of the ventilator breathing system. To achieve the
ventilator performance specified in this manual, the ventilator
breathing system must meet the specifications listed in Table 11-11.
Table 11-11: Accessory Requirements
Parameter
Compliance
Resistance
Bacteria filter
Specification
100 to 240 VAC
50/60 Hz
300 VA
PN 1076374: 14.4 V, 11.0 Ah, 163 Wh
Maximum system current draw: 11 A
Charge voltage: +16.9 V maximum
Operating time: 360 minutes (6 hours) under normal
conditions
Specification
For volume accuracy requirements, the patient circuit
compliance should be 0.98 mL/cmH
For all other performance requirements except
volume accuracy, the maximum compliance of the
circuit can be up to 2.8 mL/cmH
Maximum resistance of the breathing circuit and at-
tachments: 2.9 cmH
O at 60 L/min.
2
Dead Space: 66 ml
Bacterial/Viral Filter Efficiency: >99.99%
Resistance (@ 0.5 L/s): 0.7 cmH
Male Connector: 15 mm I.D./22 mm O.D.
Female Connector: 22 mm I.D.
O.
2
O.
2
O (hPa)/L/s
2