Respironics V60 User Manual page 116

Ventilator
Table of Contents

Advertisement

Alarms, messages, and troubleshooting
Table 9-3: Alarm and other messages: summary and troubleshooting (continued)
Message
Description
Cannot Reach Target
Displays when HFT (high
Flow
flow therapy) is active. In-
dicates that flow target is
not achieved.
Check Vent:
See Table 9-4 on page 9-13
description of failure
High Inspiratory Pres-
Measured inspiratory pres-
sure
sure is greater than the
HIP setting, and the venti-
lator cycles into exhala-
tion. Autoresets after a
complete inspiration with-
out the alarm condition.
High O
Supply Pres-
O
2
2
sure
er than 92 psig, so O
richment ends. Autoresets
when O
falls below 87 psig.
High Rate
Measured respiratory rate
is greater than the Hi Rate
setting. Escalates to a
high-priority alarm if the
alarm condition persists
for more than 60 sec.
High Tidal Volume
Measured estimated tidal
volume is greater than the
Hi V
a high-priority alarm if the
alarm condition persists
for more than 60 sec.
Low Inspiratory Pres-
Measured inspiratory pres-
sure
sure is less than the LIP
setting.
9-8
Corrective Action
Check the patient. Check
that the high flow nasal
cannula size is appropri-
ate for the flow setting.
Check that an occlusive
interface is NOT in use (a
cannula fully sealed with-
in the nares, an NIV mask
or direct connection to an
ETT/trach). Check for an
occlusion, kink or liquid in
the patient circuit.
Check the patient. Con-
firm ventilator and alarm
settings are appropriate. If
problem persists, provide
alternative ventilation.
Have the ventilator ser-
viced.
inlet pressure is great-
Check the patient. If prob-
en-
lem persists, provide alter-
2
native ventilation. Have
supply pressure
the ventilator serviced.
2
Check the patient. Con-
firm ventilator and alarm
settings are appropriate. If
problem persists, provide
alternative ventilation.
Have the ventilator ser-
viced.
Check the patient. Check
for large leaks. Confirm
setting. Escalates to
ventilator and alarm set-
T
tings are appropriate. If
problem persists, provide
alternative ventilation.
Have the ventilator ser-
viced.
Check the patient. Con-
firm ventilator and alarm
settings are appropriate. If
problem persists, provide
alternative ventilation.
Have the ventilator ser-
viced.
Priority
Manually
Autores
type (ID)
resettable
ettable
Low
No
Yes
(53)
High
Yes
Yes
(45)
High
No
Yes
(49)
Low/
Yes
Yes
High
(59)
Low/
Yes
Yes
High
(58)
High
Yes
Yes
(46)
Silence
able
Yes
Yes
Yes
Yes
Yes
Yes