Technical specifications
11-10
Table 11-13: Other specifications (continued)
Parameter
High flow therapy
(3.00 software and above, and
V60 Plus)
Pressure range
Dynamic pressure regulation
Start-up time
Triggering, cycling, and leak tol-
erance
Inspiratory and expiratory pres-
sure drop following equipment
failure: measured at patient
connection, when the recom-
mended breathing system is in
use.
Maximum time required for the
O
concentration in the deliv-
2
ered volume to change from a
volume fraction of 21% to 90%
using the worst-case ventilator
breathing system or using the
maximum internal volume venti-
lator breathing system.
The 10 s average input flow re-
quired by the ventilator for each
gas at a pressure of 280 kPa
The maximum transient input
flow averaged for 3 s required by
the ventilator for each gas at a
pressure of 280 kPa
*
Operational acoustics
ISO/FDIS 80601-2-12:2010
201.9.6.2.1.101
* In accordance with 3rd Edition testing methods
Specification
10 to 80 L/min BTPS
NOTE:
The maximum deliverable flow rate varies
based on the orifice size of the nasal
cannula and on the patient's nasal passage
resistance.
4 to 40 cmH
O
2
± (2 cmH
O + 4% of target)
2
NOTE:
Negative (subatmospheric) pressure settings
are not available.
Ready to ventilate 9 seconds after power on
As per the Digital Auto-Trak Sensitivity algorithms (see
"Auto-Trak Sensitivity" on page 4-3)
< 2.0 cmH
O (at 60 LPM)
2
< 1.0 cmH
O (at 30 LPM)
2
The ventilator adjusts O
within one breath.
2
FiO
within the gas delivery system and entire breath-
2
ing circuit adjusts at the following rate:
Up to 18 seconds for delivered volume of 500 mL
Up to 17 seconds for delivered volume of 200 mL
107 SLPM
130 SLPM
Average noise level is less than 45 dB(A) when mea-
sured 1 m from the ventilator