Pneumatic
specifications
Table 11-9: Pneumatic specifications
Parameter
High-pressure oxygen supply
High-pressure oxygen supply
(using V60/V60 Plus manifold)
Air supply
Inspiratory outlet
(to patient port)
Maximum limited pressure
(P
)
LIMmax
Maximum working pressure range
(P
)
Wmax
Subatmospheric pressures
generated during exhalation
Technical specifications
Specification
Connector: DISS male, DISS female, NIST
Pressure: 2.76 to 6.00 bar / 276 to 600 kPa / 40 to
87 psig
Flow: 175 SLPM
Connector: SIS
Pressure: 3.31 to 6.00 bar / 331 to 600 kPa / 48 to
87 psig
Flow: 175 SLPM
Connector: DISS male, DISS female, NIST
Pressure: 3.10 to 6.00 bar / 310 to 600 kPa / 45 to
87 psig
Flow: 175 SLPM
Connector: SIS
Pressure: 3.66 to 6.00 bar / 366 to 600 kPa / 53 to
87 psig
Flow: 175 SLPM
Integrated blower
Connector: ISO 15 mm female/22 mm male conical
64 cmH
O
2
5 to 50 cmH
O, ensured by High Inspiratory Pressure
2
(HIP) alarm limit
None