Alarms; Menu Window Settings - Respironics V60 User Manual

Ventilator
Table of Contents

Advertisement

Technical specifications

Alarms

Menu window
settings
11-4
Table 11-4 lists the adjustable alarm ranges and resolutions. Table 9-3 on
page 9-7 describes other, nonadjustable alarms.
Table 11-4: Adjustable alarm ranges and resolutions
Parameter
Range
Hi Rate (High Rate
5 to 90 BPM
Alarm)
Lo Rate (Low Rate
1 to 89 BPM
Alarm)
Hi V
(High Tidal Vol-
200 to 3500 mL
T
ume Alarm)
BTPS
Lo V
(Low Tidal Vol-
OFF, 5 to 1500 mL
T
ume Alarm)
BTPS
HIP (High Inspiratory
5 to 50 cmH
Pressure Alarm)
LIP (Low Inspiratory
OFF, 1 to 40 cmH
Pressure Alarm)
V
Lo
(Low Minute
OFF, 0.1 to
E
Ventilation Alarm)
99.0 L/min BTPS
LIP T (Low Inspiratory
5 to 60 sec
Pressure Delay Time
Alarm)
Table 11-5: Menu window settings and ranges
Parameter
Brightness
Loudness
Mask/ET Selection
Exhalation Port Selection
Screen Lock
Auto-Trak+ (optional)
Resolution
1 BPM
1 BPM
5 mL
5 mL
O
1 cmH
O
2
2
O
1 cmH
O
2
2
0.1 L/min
1 sec
Range
1 to 5
1 to 10
ET/Trach, 1, 2, 3, 4, Other
• DEP (Philips Respironics Disposable Exhalation
Port
• Whisper Swivel (Philips Respironics Whisper
Swivel),
• PEV (Philips Respironics Plateau Exhalation Valve),
• Other (Other Exhalation Port),
• None (No circuit exhalation port)
Off, On
Trigger: Normal, +1 to +7.
E-Cycle: -2 to -1, Normal, +1 to +6
Factory default
30 BPM
10 BPM
2500 mL
OFF
50 cmH
O
2
OFF
OFF
20 secs