EDAN SE-1010 User Manual page 9

Hide thumbs Also See for SE-1010:
Table of Contents

Advertisement

SE-1010 PC ECG User Manual
16. SHOCK HAZARD - Don't connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlet supplying the system.
17. Do not connect any equipment or accessories that are not approved by the
manufacturer or that are not IEC/EN 60601-1-1 approved to the system. The
operation or use of non-approved equipment or accessories with the system is not
tested or supported, and system operation and safety are not guaranteed.
18. Any non-medical equipment (such as the external printer) is not allowed to be used
within the patient vicinity (1.5m/6ft.).
19. Do not exceed the maximum permitted load when using the multiple portable
socket-outlet(s) to supply the system.
20. Multiple portable socket-outlets shall not be placed on the floor.
21. Do not use the additional multiple portable socket-outlet or extension cord in the
medical electrical system, unless it's specified as part of the system by manufacturer.
And the multiple portable socket-outlets provided with the system shall only be used
for supplying power to equipment which is intended to form part of the system.
22. Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing
equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configuration
shall comply with the valid version of the standard IEC/EN 60601-1-1. Therefore
anybody, who connects additional equipment to the signal input or output connector to
configure a medical system, must make sure that it complies with the requirements of
the valid version of the system standard IEC/EN 60601-1-1. If in doubt, consult our
technical service department or your local distributor.
23. Connecting any accessory (such as external printer) or other device (such as the
computer) to this electrocardiograph makes a medical system. In that case, additional
safety measures should be taken during installation of the system, and the system
shall provide:
a) Within the patient environment, a level of safety comparable to that provided by
medical electrical equipment complying with IEC/EN 60601-1, and
b) Outside the patient environment, the level of safety appropriate for non-medical
electrical equipment complying with other IEC or ISO safety standards.
24. All the accessories connected to system must be installed outside the patient vicinity,
if they do not meet the requirement of IEC/EN 60601-1.
WARNING
- 3 -
Safety Guidance

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Eda082Eda044

Table of Contents