Guidance and manufacture's declaration - electromagnetic immunity -
Guidance and manufacture's declaration - electromagnetic immunity
SE-1010 PC ECG is intended for use in the electromagnetic environment specified below. The
customer or the user of SE-1010 PC ECG should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical
fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
Voltage dips, short
interruptions
and
voltage variations
on power supply
input lines
for all EQUIPMENT and SYSTEMS
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV line to line
±2 kV line to groud
3A/m
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
Compliance level
±6 kV contact
±8 kV air
±2 kV contact
(only for system
with tablet PC)
±2 kV air
(only for system
with tablet PC)
Not applicable
Not applicable
3A/m
Not applicable
- 75 -
EMC Information
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic
material,
relative humidity should be
at least 30%.
Mains
power
should be that of a typical
commercial
or
hospital
environment.
Mains
power
should be that of a typical
commercial
or
hospital
environment.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location
in
a
commercial
or
hospital
environment.
Mains
power
should be that of a typical
commercial
or
hospital
environment. If the user of
SE-1010 PC ECG requires
the
quality
quality
typical
quality