SE-1010 PC ECG User Manual
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which SE-1010 PC ECG is
used exceeds the applicable RF compliance level above, SE-1010 PC ECG should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating SE-1010 PC ECG.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Field
strengths
transmitters, as determined by an
electromagnetic site survey,
less than the compliance level in each
frequency range.
Interference may occur in the vicinity
of
equipment
following symbol:
- 77 -
EMC Information
from
fixed
RF
a
should be
b
marked
with
the