Attaching The Reusable Electrodes; Attaching The Limb Electrodes - EDAN SE-1010 User Manual

Hide thumbs Also See for SE-1010:
Table of Contents

Advertisement

SE-1010 PC ECG User Manual
European
American
Label
C1
C2
C3
C4
C5
C6
L
R
F
N

3.3.2 Attaching the Reusable Electrodes

3.3.2.1 Attaching the Limb Electrodes

Connecting to a Lead Wire
Electrode Placement
Label
V1
Fourth intercostal space at the right border of the sternum
V2
Fourth intercostal space at the left border of the sternum
V3
Fifth rib between C2 and C4
V4
Fifth intercostal space on the left midclavicular line
V5
Left anterior axillary line at the horizontal level of C4
V6
Left midaxillary line at the horizontal level of C4
LA
Right arm/Right deltoid
RA
Left arm/Left deltoid
LL
Right leg/Upper leg as close to torso as possible
RL
Left leg/Upper leg as close to torso as possible
Clamp
Limb Electrode Connection:
1) Ensure that the electrodes are clean;
2) Clean the electrode area which is a short distance
3) Daub the electrode area on the limb with gel evenly;
4) Place a small amount of gel on the metal part of the
5) Connect the electrode to the limb, and make sure that
6)
Limb Electrode
above the ankle or the wrist with 75% alcohol;
limb electrode clamp;
the metal part is placed on the electrode area above
the ankle or the wrist;
Attach all limb electrodes in the same way.