SE-1010 PC ECG User Manual
Item
Diagnosis
Rhythm Lead Set
Print Format
Description
Select the diagnosis item displayed on the preview screen and in the
report printed out.
NOTE: Position Mark should be selected together with Average
template, because the position mark is only used to mark
the position of ECG waves in the average template.
Set the rhythm lead to one of 12 standard leads: І, ІІ, ІІІ, aVR, aVL, aVF,
V1, V2, V3, V4, V5, or V6.
When the printing mode is set to 3×4+1 or 6×2+1, the rhythm lead
selected in the Rhythm1 list box will be printed out.
When the printing mode is set to 3×4+3, 3 rhythm leads selected in the
Rhythm1, Rhythm2 and Rhythm3 list boxes will be printed out.
♦ Set Orient to landscape or portrait.
♦ Set Sequence to sequential or synchronous.
Select sequential, the lead group is printed one by one in a certain
sequence. The start time of a lead group is just the end time of the
previous lead group.
synchronous, all leads are printed simultaneously. The start
Select
time of each group is the same.
♦ Set Paper Size to A4 or Letter.
♦ Set Report Hint to Report confirmed by: or Unconfirmed Report.
♦ Set Baseline Adjustment to OFF, Horizontal or Auto.
♦ Select Auto gain change, and the gain will be changed automatically
and Baseline Adjustment automatically.
♦ Select Color, the background grid of report will be printed in color.
NOTE: If the printing color is set to color, but a black-and-white
printer is used, the report printed will be illegible.
♦ Select Background grid, and the background grid will be printed in
the report.