SE-1010 PC ECG User Manual
CMRR
Pacemaker Detection
Amplitude
Width
Sampling Frequency
NOTE: Test the accuracy of input signal reproduction according to the methods
described in clause 4.2.7.2 in ANSI/AAMI EC11:1991/(R) 2001/(R) 2007, and
the result complies with clause 3.2.7.2 in ANSI/AAMI EC11:1991/(R) 2001/(R)
2007.
DP12 ECG sampling box≥110 dB
DX12 transmitter≥100 dB
DP12 ECG sampling box: ±2 to ±700 mV
DX12 transmitter: ±2 to ±500 mV
0.1 to 2.0 ms
DP12 ECG sampling box:10,000 Hz / rhythm lead
DX12 transmitter: 10,000 Hz / lead
- 73 -
Technical Specifications