Keysight N4373E Lightwave Component Analyzer Setup Guide
To prevent electrical shock, disconnect the instrument from mains
before cleaning. Use a dry cloth or one slightly dampened with water to
clean the external case parts. Do not attempt to clean internally.
Using controls or adjustments or performing procedures other than
those specified in the documentation supplied with your equipment can
result in hazardous radiation exposure.
This product complies with over-voltage Category II and Pollution
Degree 2.
Ventilation requirements: When installing the product in a cabinet, the
convection into and out of the product must not be restricted. The
ambient temperature (outside the cabinet) must be less than the
maximum operating temperature of the product by 4 °C for every
100 watts dissipated in the cabinet. If the total power dissipated in the
cabinet is greater than 800 watts, then forced convection must be used.
Install the instrument so that the power switch is readily identifiable and
is easily reached by the operator. This is the instrument disconnecting
device. It disconnects the mains circuit from the mains supply before
other parts of the instrument. Alternatively, an externally installed switch
or circuit breaker (which is readily identifiable and is easily reached by
the operator) may be used as a disconnecting device.
Always use the three-prong AC power cord supplied with this
instrument. Failure to ensure adequate earth grounding by not using this
cord may cause instrument damage.