Keysight Technologies N4373E Setup Manual page 31

Lightwave component analyzer
Table of Contents

Advertisement

CAUTION
Keysight N4373E Lightwave Component Analyzer Setup Guide
Before you move the system or return it to a service center, make sure
NOTE
you
• put the rigid cables back into the horizontal position and
• reattach the cable guard.
3
Remove the dust cover from Port B on the optical test set.
4
Attach the female to female adapter (part number 85058-60121) to
Port B.
5
Screw the adapter finger-tight to Port B without rotating it.
6
Holding the adapter with the spanner wrench from the calibration kit
(part number 08513-20014), use the 20mm torque wrench from the
calibration kit (0.90 N.m/8 lb.in, part number 8710-1764) to tighten
the adapter.
For information on handling, calibrating or cleaning RF connectors,
NOTE
please refer to the "User's and Service Guide Keysight Technologies
N4697E/F 1.85 mm Flexible Cables for Test Ports", available on the
Keysight website.
For further information on the Adapter, please refer to the "User's and
Service Guide Keysight Technologies 85130H Rugged 1.85 mm to
1.85 mm Adapter Kit".
Always rotate the nut and never the female to female adapter.
Rotating the adapter can damage the electrical interface of the adapter
and the connector on the optical test set.
Setting Up the Instrument
2
31