Setting Up the Instrument
32
For information on handling using and maintaining the flexible test port
NOTE
cables, please refer to the Operating and Service Manual, Keysight
N4697E/F NMD-1.85 mm -f- to 1.85 mm Flexible Test Port Return
Cables (part no. N4697-90001).
7
Remove the dust cover from Port 2 of the 2-port network analyzer, or
from Port 4 of the 4-port network analyzer. Remove the dust cover
from both ends of the flexible test port cable (part number
N4697-60030 or 85133-60043).
8
Connect the test port cable finger-tight to Port B on the optical test set
without rotating the female to female adapter. Connect the other end
to Port 2 of the 2-port network analyzer or, respectively, to Port 4 of
the 4-port network analyzer.
9
Holding the adapter with the spanner wrench, use the 20 mm torque
wrench from the calibration kit to tighten both connections.
Keysight N4373E Lightwave Component Analyzer Setup Guide