Cleaning and maintenance
66
Never drop spacer discs
n
into the aluminum housing.
Check for any spacer discs
n
adhering to the connecting
rod.
Keep the spacer discs. It is
n
essential to reinsert the
same number of spacer
discs.
6. Assemble the diaphragm
assembly and ensure its
correct positioning on the
square head.
20901367_EN-US_PC3001 VARIO select Serie_V1.3_100720
5. Disassemble the assembly
and take a new di-
aphragm; service kit
MD 1C.
Ensure that the diaphragm
n
is inserted correctly, with
the coated, light-colored
side facing upwards.
Pay special attention to
n
correct positioning on the
square head.
7. Use the correct number of
spacer discs.