Calibrate for instal-
lation conditions
-> Example
Drawing showing
minimum distances
in the laboratory fur-
niture
20901367_EN-US_PC3001 VARIO select Serie_V1.3_100720

4.2 Installation

NOTICE
Condensate can damage the electronics.
A large temperature difference between the storage location
and the installation location can cause condensation.
ð After goods receipt or storage, allow your vacuum device to
acclimatize for at least 3-4 hours before initial use.
Check installation conditions
The device is acclimatized.
n
Ambient conditions have been observed and are within the
n
limitation of use.
The pump must have a stable and secure base without addi-
n
tional mechanical contact apart from the pump feet.
Vacuum pump installation
ð Place the vacuum pump on a stable, non-vibrating, level
surface.
ð When installing in lab furniture, maintain a minimum dis-
tance of 5 cm (2 in) to adjacent objects or surfaces.
ð Prevent heat accumulation and ensure sufficient air circula-
tion, especially in closed housings.
Installation and connection