vacuubrand PC 3001 VARIO select Instructions For Use Manual page 57

Chemistry pumping unit
Hide thumbs Also See for PC 3001 VARIO select:
Table of Contents

Advertisement

-> Example
Remove the sensor
-> Example
Clean the measure-
ment chamber and
the venting valve
20901367_EN-US_PC3001 VARIO select Serie_V1.3_100720
4. Close the hose nozzle of
the chiller and carefully
turn the pumping unit up-
side down.
6. Undo the screws, hex key
size 3
Clean the sensor
1. Use a pipette to fill a small
amount of solvent,
e.g., benzene, into the
openings.
3. Repeat until the solvent is free from contaminants.
4. Let the interior of the sensor air dry or dry under vacuum.
Cleaning and maintenance
5. Open the union nut on the
sensor, open-end wrench
SW17, and pull off the
molded hose.
7. Disconnect the
VACUU·BUS connector at
the bottom and remove
the sensor.
2. Leave the solvent to act
for a few minutes before
pouring it off.
57