Coolant connection
IN = feed line
EX = outlet
-> Example
Coolant connection
at the EK or IK
20901367_EN-US_PC3001 VARIO select Serie_V1.3_100720

4.3.3 Coolant connection at the condenser

A vapor condenser (EK) has a connection for coolant. For ex-
ample, water or liquid in a chiller circuit is suitable for cooling.
In a closed, in-house coolant circuit, the pressure should be
n
limited to 3 bar (44 psi).
A coolant valve may only be installed in the feed line; the
n
coolant drain must be clear and non-pressurized.
Connect the coolant
1. Fasten both hose nozzles (a) with union nuts (b) on the
condenser as shown.
2. Fasten the hoses for the coolant on the condenser as
shown in the diagram:
IN = inlet
EX = outlet
3. Secure the hoses, e.g., with hose clips.
5 Also applies to the inlet condenser IK
Installation and connection