Appendix | Measurement on Narrow Targets
A 3

Measurement on Narrow Targets

50 %
45 %
40 %
35 %
30 %
25 %
20 %
15 %
10 %
5 %
0 %
0
0.5
1
1.5
Target dispacement perpendicular to the sensor axis [mm]
Fig. 19 Example measuring range deviation with a
sensor distance of 10% of the measuring range
50 %
45 %
40 %
35 %
30 %
3 mm
25 %
4 mm
20 %
15 %
10 %
5 %
0 %
0
0.5
1
1.5
Target dispacement perpendicular to the sensor axis [mm]
Fig. 21 Example measuring range deviation with a
sensor distance of 100% of the measuring range
capaNCDT 61x4
50 %
45 %
40 %
35 %
30 %
25 %
20 %
3 mm
15 %
4 mm
10 %
6 mm
5 %
8 mm
0 %
2
2.5
3
Fig. 20 Example measuring range deviation with a sen-
sor distance of 50% of the measuring range
6 mm
8 mm
2
2.5
3
Fig. 22 Signal change when thin targets are shifted
transversely to the measurement direction
3 mm
4 mm
6 mm
0
0.5
1
1.5
Target dispacement perpendicular to the sensor axis [mm]
y
y >8 mm
8 mm
2
2.5
3
z
z constant
x
Movement
in x-direction