Terumo CDI 550 Operator's Manual page 105

Blood parameter monitoring system
Table of Contents

Advertisement

Error code
(Recoverable)
F033
F034
F035
F036
F038
F039
F050
F070
9-20
chapter 9
Description
Display CDI error
message at start up.
Display CDI error
message at start up.
Display CDI error
message at start up.
Display CDI error
message at start up.
Display CDI error
message at start up.
Display CDI error
message at start up.
Display CDI error
message at start up.
Display CDI error
message at start up.
Cause
Arterial PC card error.
Venous PC card error.
HSAT Probe error.
HSAT Servo card error.
AUX board failure, or
software update.
RTC (real time clock)
checksum error.
SBC board failure.
Barometric pressure
related
Arterial PC card Rev
code error, or card is
not plugged in.
Recommendation
Allow user to ignore error
message. If ignored,
arterial BPM is disabled.
Venous BPM, if installed,
and H/Sat Probe will still
be available for use.
Allow user to ignore error
message. If ignored,
venous BPM is disabled.
Arterial BPM and H/Sat
Probe will still be
available for use.
Allow user to ignore error
message. If ignored,
H/Sat probe is disabled.
Venous BPM, if installed,
and arterial BPM will still
be available for use.
Allow user to ignore error
message. If ignored,
H/Sat probe is disabled.
Venous BPM, if installed,
and arterial BPM will still
be available for use.
Allow user to ignore the
error message. If ignored,
all setup parameters will
be set back to Factory
Defaults.
Allow user to ignore the
error message. If ignored,
date settings will be set
back to 1/1/1997, time
will be set to 12:00. All
the other functions are
available.
Allow user to ignore error
message. If ignored,
760mmHg will be used
for calibration.
Allow user to ignore error
message. If ignored,
Arterial BPM is disabled.
Venous BPM, if installed,
and H/Sat Probe will still
be available for use.