Terumo CDI 550 Operator's Manual page 73

Blood parameter monitoring system
Table of Contents

Advertisement

7-4
chapter 7
Caution
Ensure that the stored and recalled temperature mode are the same as
the laboratory sample to ensure proper calibration. Calibrating to different
temperature modes will compromise the system accuracy limits, as
described in Appendix B.
4
When the lab value is received, press the "recall" key.
The recalibration window opens (as shown in the next figure). The potassium
value highlights (black background) automatically when you press recall for the
first time.
recalibration14:37
7.45
PCO 2
36
Enter new values and press the OK ( ) key ...
PO 2
pH
PCO2 PO2
124
SO2
Stored
7.45
37
124
98
Values
Adjusted
7.45
37
124
98
37.1
Actual
Values
HCO
BE
calc
SO
3
High
High
22
-23
98
K+
CAL
VO
4.3
2
mmol/l
setup
calibration
Use the + and – keys to change the highlighted value.
5
Change the potassium value, to match the lab value, by using the + and −
or
and
keys.
6
Change the other blood parameter values to match the lab value, by us-
ing the + and – or
7
Press √ when you've finished adjusting all values.
Or, press X to cancel the adjustment and leave the "recall" key active.
Warning
Measured values prior to initial in vivo calibration may not be
accurate. Do not use values prior to initial in vivo calibration for
patient management. Failure to perform a proper set-up, full two-
point tonometered gas calibration, and initial in vivo calibration may
result in compromised system performance that does not meet system
accuracy limits found in Appendix B. To perform additional in vivo
recalibrations see page 7 –9 , "Monitoring a Patient: Recalibrating
During a Procedure ."
Note: If you press √ without changing any values, the stored potassium value be-
comes the adjusted value and is used to set the potassium slope. Upon returning
to operate, the "CAL" indicator will disappear and the K
The other stored blood parameter values will also become the adjusted values. Upon
returning to operate, the values will no longer appear dimmed.
13:26
9-25-18
7.35
PCO 2
42
High
High
mmHg
mmHg
PO 2
High
K+
pH
PCO2 PO2 SO2 HCT Hgb
39
High
4.3
7.35
42
39
78
20
6.7
mmHg
mmHg
Temp
4.3
7.35
42
39
78
20
6.7
36.4
Actual
c
c
HCT
Hgb
calc
SO
2
2
High
High
High
High
20
6.7
78
%
%
%
g/dl
Q
92
4.3
High
High
ml/min
l/min
standby
operate
numeric
and
keys and and keys.
@
°
C
pH-stat
store
set Q
mark
When the recalibration
feed
screen first opens, the
potassium value is
highlighted.
5000705
+
value will appear normal.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents