Command List.
The following is a table of major Command values:
Value
Name
' A '
Attention
'K'
Acknowl-
edge
'N'
No
Acknowl-
edge
'E'
Error
'S'
Set
'G'
Get
Direction
From CDI
Information sent out in response to a request or a timed
System 550
action, such as calibration data, or sending out a line of
analyte data during Operate mode.
From CDI
Acknowledge a request by echoing back the Command
System 550
and SubCommand, along with the requested information
in the Data portion of the packet. Acknowledge packets
are sent back to the requesting device (usually the CDI
System 550) to both confirm that the request came
through successfully, and to provide any data associated
with the request.
From CDI
Indicates a problem with packet communications. See
System 550
the SubCommand list of NAK conditions for more details.
From CDI
Indicates an error condition in the CDI System 550. The
System 550
data portion of the packet will contain the error code,
which will be the same as seen on the CDI System 550
screen.
To CDI Sys-
Sent to the CDI System 550 to set certain parameters,
tem 550
such as Arterial Flow or certain toggles.
To CDI Sys-
Sent to the CDI System 550 to request information about
tem 550
the state of an element in the system, such as the time or
the operate state.
Description