Terumo CDI 550 Operator's Manual page 52

Blood parameter monitoring system
Table of Contents

Advertisement

The following table lists the possible settings for the parameters on the serial port
configuration screen.
Output Type
Possible settings are "ASCII Output" , "Terumo® Systems," or
"Packet Mode" .
Data Bits
Possible settings are "8 data/1 stop" or "7 data/2 stop."
Parity
Possible settings are None, Even, or Odd.
Baud Rate
Possible settings are 1200, 2400, 4800, 9600, 19200, or
38400.
Output Interval
Possible values are 0, 0.1, 0.5, or 1-10 minutes (in 1 minute
increments). Determines how often data will be sent to an
external device via the serial port. If this option is set to 0, the
CDI System 550 sends data on demand only.
Note: The data bit, parity, and baud rate selections will depend on the specific receiv-
ing device.
Note: The serial port settings are applicable to the data output port only (top 9-pin
connector on rear panel). The pump interface port, directly beneath the data output
port, is setup through the "Pump Type" and " Source" selection in screen 5.
Note: See Appendix C for more information on the use and configuration of the serial
port.
When you've finished adjusting all the system settings, you can switch from setup
mode to a different mode. Here's how:
1
Press the system mode select key to activate the system map.
2
Use the and keys to select the mode you want.
3
Confirm your choice of modes by pressing the √ key.
choosing system settings
4-13