Error code
(Recoverable)
F233
F234
F0A0
F0A1
F0A2
F0A3
9-22
chapter 9
Description
Displayed in EEPROM
dump.
Displayed in EEPROM
dump.
Display CDI error
message at start up.
Display CDI error
message at start up.
Display CDI error
message at start up.
Display CDI error
message at start up.
Cause
Arterial BPM Probe
EEPROM failure
Venous BPM Probe
EEPROM failure
Arterial BPM Probe
ADC failure.
Arterial BPM Probe
thermistor failure.
Venous BPM ADC
failure.
Venous BPM
thermistor failure.
Recommendation
Allow user to ignore error
message. If ignored,
arterial BPM is disabled.
All other functions are
available.
Allow user to ignore error
message. If ignored,
venous BPM is disabled.
All other functions are
available.
Allow user to ignore error
message. If ignored,
arterial BPM is disabled.
All other functions are
available.
Allow user to ignore error
message. If ignored,
arterial BPM is disabled.
All other functions are
available.
Allow user to ignore error
message. If ignored,
venous BPM is disabled.
All other functions are
available.
Allow user to ignore error
message. If ignored,
venous BPM is disabled.
All other functions are
available.