Message/Problem
Calibrator failure
detected: _________
Contact Technical
Service.
Calibration (or verifica-
tion) process stopped.
Calibrator not
connected.
Check connection.
Meaning and Corrective Action
This message refers to a detected calibrator hard-
ware failure or communication failure with the moni-
tor. An identifying error code appears in the blank.
Record this error code for later reporting to Terumo
Technical Service. At this time, you have the option
to re-initiate the calibration procedure. If the error
code persists, re-attempt calibration with a new cali-
brator if available. Success of the calibration process
with another calibrator would indicate a hardware
failure with the first calibrator. Press the X (cancel)
key to exit calibration and return to standby. If no
alternate calibrator is available, the system will allow
continued use with application of the system default
calibration values. Contact your Terumo Technical
Service representative to arrange for repair.
This message occurs when the "stop" soft key is
pressed during calibration (or verification). Pressing
the √ (OK) key will restart calibration (or verification);
pressing the X (Cancel) key will exit calibration and
will put the monitor in standby mode.
The monitor did not detect the presence of a calibra-
tor when it tried to initiate the calibration sequence.
This occurs if the calibrator cable is not connected
or is loosely connected. Check the connection and
press the √ (OK) key to try calibration again. If the
calibrator is still not detected after checking the con-
nector, contact your Terumo Technical representative
to arrange repair of the system. Use of the moni-
tor can continue without the calibration function by
pressing the X key (cancel) to exit calibration. You
may try another calibrator if one is available.
Note: If the calibrator has become disconnected
during calibration, and reconnection has successfully
occurred, the calibration will start over again.