Message/Problem
Meaning and Corrective Action
Arterial (or Venous
This message refers to a detected module hardware fail-
or H/S) module
ure or communications failure with the monitor. This may
failure: ________
occur due to sudden component failure or physical dam-
age to the module, monitor or cable. An identifying error
Measurements
code appears in the blank. Record this error code for later
from module will
reporting to Terumo Technical Service. You may continue
be discontinued.
using the system although measurements from this mod-
ule will be discontinued and will no longer be displayed.
Contact Terumo
Contact your Terumo Technical Service representative to
Technical Service.
arrange for repair.
Monitor failure
This message refers to a detected monitor hardware fail-
detected:
ure. This may occur due to sudden physical damage to
________
monitor, component failure, or failure of all modules. An
identifying error code appears in the blank. Record this
Contact Terumo
error code for later reporting to Terumo Technical Ser-
Technical Service.
vice. You will not be able to continue use of the system.
Contact your Terumo Technical Service representative to
arrange for repair.
troubleshooting
9-11