Terumo CDI 550 Operator's Manual page 157

Blood parameter monitoring system
Table of Contents

Advertisement

C-26
appendix c
B − Unit descriptor − hex-encoded ASCII value indicating the units for this analyte.
Use the table below for definitions:
No Unit
Degrees C
mmHg
kPa
Percent
Milliequivalents/Liter
Grams/Deciliter
Millimoles/Liter
Milliliters/Minute
Milliliters/Minute/Meter²
Liters/Minute
C – Analyte Status Bitmap – hex-encoded ASCII bitmap value indicating various sta-
tus information. Since it is only a single character, there are only 4 bits of status.
Bit 0 (LSB)
Bit 1
Bit 2
Bit 3 (MSB)
This means that if the status value is in the range 0x8 – 0xF, the analyte is unavail-
able due to a configuration selection.
D..D – Analyte Value – The value of the analyte will be displayed in one of three pos-
sible formats:
Integer 2 or 3 digit
Floating point x.xx or xx.x
Dashes of the above formats
Examples:
Arterial O
of 54 in mmHg at Actual Temperature with Low Alarm status:
2
Potassium of 5.4 in mEq/L at 37 degree value with no alarms:
Hematocrit in Percent unavailable while measuring in Actual Temperature: 104A--
Unit
0: 37 degree value, 1: Actual temperature value
0: Normal data, 1: Low alarm status
0: Normal data, 1: High alarm status
0: Normal data, 1: Analyte unavailable
Unit Code
'0'
'1'
'2'
'3'
'4'
'5'
'6'
'7'
'8'
'9'
' A '
022354
07505.4