Attaching the Sensor
The sensor can be connected to the device before or after applying the device to the patient.
The following steps apply to these Nonin sensors:
• 8000SS-WO2, 8000SM-WO2, 8000SL-WO2
• 8000AA-WO2
• 8000J-WO2
• 6000CA-WO2
NOTE: Refer to the sensor Instructions for Use for appropriate sensor sizing.
If using 8000R or 8000QR Noninsensor, use sensor adapter cable 3150I (see "Parts and
Accessories").
WARNING: Only use Nonin-branded sensors with a length of 1 meter or less.
Accuracy may degrade if sensor cable is over 1 meter in length. Using the sensor
cable adapter does not affect accuracy.
1. Insert the sensor connector into the sensor port at the top of the device (figure 12). The
Nonin logo on the sensor connector should face the front of the device.
2. Push the connector until it clicks into place.
3. The device is ready to use.