Advertisement

Next step
Check & Adjust
4.
Measure to measure and to
continue to the next screen.
Motorised instruments change
automatically to the other face.
Non-motorised instruments
guide to the other face.
The fine pointing must
be performed manu-
ally in both faces.
5.
Face II measurement
Measure to measure the same target in the other face and to cal-
culate the tilting axis error.
If the error is bigger than the predefined limit, the procedure must
be repeated. The tilting axis measurements of the current run are
then rejected and not averaged with the results from previous runs.
6.
Adjustment Status
Number of measurements: Shows the number of runs completed.
One run consists of a measurement in face I and face II.
a T-axis quality (1 σ): shows the standard deviation of the determ-
ined tilting axis error. The standard deviation can be calculated from
the second run onwards.
Measure at least two runs.
7.
Next to continue with the check & adjust procedure.
8.
Select Add another calibration loop if more runs have to be
added. Next and continue with step
OR
Select Finish the calibration & store the results to finish the cal-
ibration process. No more runs can be added later. Next to view the
adjustment results.
9.
Select Finish to accept the results. No more runs can be added
later.
OR
Select Redo to decline all measurements and to repeat all calibra-
tion runs.
IF the results are
to be stored
to be determined again
180°
008372_001
3.
THEN
Next overwrites the old tilting axis error with
the new one.
Redo rejects the new determined tilting axis
error and repeats the whole procedure.
Refer to paragraph
Combined adjustment
procedure
step-by-step.
180°
61