Check & Adjust
If Calibrate the automatic tar-
get aiming is checked and an
ATRplus is available, the adjust-
ment will include the determina-
tion of the ATRplus Hz and V
adjustment errors.
☞
Use a clean Leica
standard prism as the
target. Do not use a
360° prism.
5.
Measure to measure and to
continue to the next screen.
Non-motorised instruments
guide to the other face.
☞
The fine pointing has
to be performed
manually in both
faces.
6.
Face II measurement
Measure to measure the same target in the other face and to cal-
culate the instrument errors.
☞
If one or more errors are bigger than the predefined limits, the pro-
cedure must be repeated. All measurements of the current run are
rejected and none of them is averaged with the results from previ-
ous runs.
7.
Adjustment Status
Number of measurements: Shows the number of runs completed.
One run consists of a measurement in face I and face II.
l Component quality (1 σ): and similar lines show the standard
deviations of the determined adjustment errors. The standard devi-
ations can be calculated from the second run onwards.
Measure at least two runs.
☞
8.
Next to continue with the check & adjust procedure.
9.
Select Add another calibration loop if more runs have to be
added. Next and continue with step
OR
Select Finish the calibration & store the results to finish the cal-
ibration process. Next to view the adjustment results.
10.
Select Finish to accept the results. No more runs can be added
later.
± 9°
008371_001
180°
180°
008372_001
4.
59