Taking measurement in Single or Both Arm(s) Mode

button to start the measurement when standby according to the settings
Press the
in Part A) when measurement of one arm or both arms is selected. The device
shows all the settings and then starts counting down the Resting Time before the first
measurement. The average measurement reading is displayed and saved after the
measurements are complete.
The device switches the display for the average
Systolic Blood Pressure (SYS), Diastolic Blood
Pressure, Mean Arterial Pressure (MAP),
central Systolic Blood Pressure (cSYS), central
Diastolic Blood Pressure, and central Pulse
Pressure (cPP) automatically if central blood
pressure measurement is enabled.
If CBP measurement is enabled, the cuff pressure is held at
*
around 60 mmHg for around 10 seconds to collect sufficient
pulse waves.
If the difference of blood pressure readings between two
*
arms is more than 15mmHg for Systolic at more than three
measurements, the "IAD" icon will show with the result. The
"IAD" icon and readings of the higher arm flash.
*
Press
button during countdown to skip the countdown.
*
Press
button to cancel remaining measurements at anytime
during the measurement sequence. Display the results
(average) if available.

Viewing stored values

The device stores blood pressure values of the last measurement(s) procedure in AUTO mode.
button to reveal the average of the measurements of AUTO mode. Continue
Press the
pressing the
button to review individual measurements.
The device switches the display for values of the individual measurement (including SYS, DIA, MAP, cSYS, cDIA and cPP
*
values) if central blood pressure measurement is enabled.

Settings of the measurement

Part B) when assessing brachial-ankle Pulse Wave Velocity (baPWV) and Ankle Brachial
Index (ABI) measurement (one arm and one ankle measurement is selected).
Set the Height of User (in cm) – Press the
selected for measurement to enter setting of Height. Use the
button to scroll down the Height value. Prese
use
to standby mode.
Height range is within 120-210cm .
*
*
Only for the device version with baPWV feature.
12
button when one arm and one ankle is
button to scroll up and
button to confirm and return