Microlife WatchBP Office Vascular Instruction Manual

Microlife WatchBP Office Vascular Instruction Manual

Professional ofice blood pressure and cardiovascular screening monitor
Hide thumbs Also See for WatchBP Office Vascular:

Advertisement

Quick Links

WatchBP
Office
Vascular
Professional Ofice
Blood Pressure and
cardiovascular screening
monitor
MODE
TWIN200 VSR
Instruction Manual
EN

Advertisement

Table of Contents
loading

Summary of Contents for Microlife WatchBP Office Vascular

  • Page 1 WatchBP Office Vascular Professional Ofice Blood Pressure and cardiovascular screening monitor MODE TWIN200 VSR Instruction Manual...
  • Page 2 Indications for Use The Microlife Upper Arm Automatic Digital Blood Pressure and Cardiovascular Screening Monitor Model WatchBP Ofice Vascular (TWIN200 VSR) is a non-invasive digital blood pressure device using oscillometric technique and an upper-arm blood pressure cuff to measure systolic and diastolic blood pressures, pulse rate, mean arterial pressure (MAP) for or use in adults and pediatrics (but not neonates) with arm cuff circumference sizes ranging from 14 -52 cm.
  • Page 3: Table Of Contents

    WatchBP Office Table of Contents Product description ・Contents ..........................4 ・Model Type ..........................4 ・Upgrading the version of the device ..................4 ・Product Overview ........................5 ・Display ..........................5 Initial set up ・Attaching the power plug to the power adapter ................ 6 ・Power ON/OFF ........................
  • Page 4: Contents

    The ABI, baPWV and central blood pressure indices measurement of the device can be activated through the WatchBP Analyzer. An activation key is needed for activation, the activation key is specific for the device as it matches the ID. Please contact Microlife or the local distributor for additional information.
  • Page 5: Product Overview

    WatchBP Office Product Overview Cuff Socket Power Adapter Socket Display USB port DC 7.5V 1.5A Battery Compartment button (Memory) To enter memory record and Down function in setting button (Up) To enter SETTING and scroll up/switch button (Start/Stop) Start measurement(s) based on the AUTO or MANUAL mode in standby MODE button (Mode)
  • Page 6: Initial Set Up

    Initial set up Attaching the power plug to the power adapter Select a suitable plug attachment and attach to the power adapter as shown here. Charge the battery completely When using the device for the first time, charge the battery until the recharge indicator on the device turns green.
  • Page 7 WatchBP Office Set the Hide function - The device features a Hide function to prevent influence of blood pressure in patients due to nervousness triggered by visible blood pressure measurements. Prese button to switch ON or OFF the Hide Function. If the Hide function is ON it means that the BP values will not be displayed during the blood pressure measurement.
  • Page 8: Before Using The Device

    M and L size upper arm cuffs are included as standard accessories. For ankle Cuff Size Circumference (cm) Circumference (inch) 22-32 8.7-12.6 32-42 12.6-16.5 M size ankle cuff is included as standard accessories. Contact Microlife or its authorized distributor to purchase cuffs. Use only cuffs provided by Microlife!
  • Page 9: Fitting The Cuff Properly

    WatchBP Office Fitting the arm cuff properly 1 Measure the patient’s upper-arm circumference to select the appropriate cuff. 2 Place the cuff over the upper arm so that the air 2–3cm tube and artery mark arrow point towards the lower (¾...
  • Page 10: Turn On The Power

    Taking measurements using the WatchBP Office Vascular Turn on the power Turn on the device by pressing the button of the device. Connect the cuff(s) to the device Connect the cuffs to the device by inserting the cuff connector into the cuff connector socket.
  • Page 11 WatchBP Office 3 Set the Interval Time – between measurements - Press the button to scroll up and use button to scroll down among 15, 30, 60, 120, 180, 240, 300 seconds of Interval Time. Prese button to confirm and enter AFIB detector setting. Set the interval time will be skipped if Number of Measurements is 1.
  • Page 12: Taking Measurement In Single Or Both Arm(S) Mode

    Taking measurement in Single or Both Arm(s) Mode button to start the measurement when standby according to the settings Press the in Part A) when measurement of one arm or both arms is selected. The device shows all the settings and then starts counting down the Resting Time before the first measurement.
  • Page 13: Taking Measurement Simultaneously In One Arm & One Ankle

    WatchBP Office Taking measurement simultaneously in one Arm & one ankle (for ABI or baPWV assessment) When standby with one Arm & one ankle measurement selected, press the button to start a measurement. The device automated takes two consecutive measurements with 60 seconds resting time and 60 seconds interval time.
  • Page 14: Using Watchbp Analyzer

    1GHz CPU. 512MB Memory, 4.5GB free hard disk space, Microsoft Windows 7 SP1 / 8 / 10 Installing the Software Program The latest WatchBP Analyzer Software is available from the Microlife website. https://www.microlife.com/support/software-professional-products Double click the download installer and simply follow the instructions provided in the installation window on the computer screen.
  • Page 15: Bluetooth Connectivity

    • When the green recharge indicator is off, it means the recharge is completed. • When the green recharge indicator lashes, it means that there is a charging error. Make sure that the correct Mains Adapter is used. If the condition persists, contact Microlife or the local distributor.
  • Page 16: Using A Power Adapter

    Using a power adapter Only use the Mains Adapter supplied with the device to recharge the device. 1) Plug the adapter cable into the power socket of the device. 2) Plug the adapter plug into the wall socket. DC 7.5V 1.5A The battery will be recharged if the device DC 7.5V 1.5A is attached to an AC power source.
  • Page 17: Device Cleaning And Disinfecting

    Accuracy test We recommend the device to be tested for accuracy every 2 years or after mechanical impact (e.g. Being dropped). Please contact Microlife to arrange an accuracy test. Disposal Batteries and electronic instruments must be disposed of in accordance with...
  • Page 18: Error Messages And Troubleshooting

    Error messages and Troubleshooting If an error occurs during measurement, the measurement is interrupted and an error message «Er» is displayed. Error Description Potential cause and remedy The pulse signals on the cuff are too weak. Reposition the cuff "Er 1" Signal too weak and repeat the measurement.
  • Page 19 Re-position the cuff and repeat the “single fault measurement. Replace the batteries if necessary. If the error condition” persists, contact Microlife or the local distributor. Possible hardware fault. Try again. If the error persists, contact Flash memory “Er A”...
  • Page 20: Technical Specifications

    • Device fulills the stipulations of the standard IEC 60601-1-2. Electromagnetic Compatibility: The stipulations of the EU Directive 93/42/EEC for Medical Devices Class IIa have been fulilled. Type BF applied part Microlife reserves the right to alter technical speciications without prior written notice.
  • Page 21 WatchBP Office Guarantee Card This device is covered by a two-year guarantee from the date of purchase. This guarantee is valid only on presentation of the guarantee card completed by the owner conirming date of purchase or purchase receipt. Batteries and wearing parts are not covered by this guarantee.
  • Page 22: Reference

    Hypertens Res, 2011. 34(7): p. 825-30. 2. Cheng, H.M., et al., Measurement accuracy of a stand-alone oscillometric central blood pressure monitor: a validation report for Microlife WatchBP Ofice Central. Am J Hypertens, 2013. 26(1): p. 42-50.
  • Page 23 WatchBP Office Federal Communications Commission (FCC) Statement Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including...
  • Page 24 Tel. +886 2 8797 1288 Fax +886 2 8797 1283 Email: watchbp@microlife.com.tw www.watchbp.com North / Central / South America Microlife USA, Inc. 1617 Gulf to Bay Blvd 2nd Floor, Suite A Clearwater, FL 33755, USA Tel. +1 727 442 5353 Fax +1 727 442 5377 Email: msa@microlifeusa.com...

This manual is also suitable for:

Twin200 vsr

Table of Contents