Microlife WatchBP Office Targe Instruction Manual

Office blood pressure measurement device with built-in blood pressure targets

Advertisement

Quick Links

Office blood pressure
measurement device
with built-in
Blood Pressure targets.
Instruction Manual
1
EN

Advertisement

Table of Contents
loading

Summary of Contents for Microlife WatchBP Office Targe

  • Page 1 Office blood pressure measurement device with built-in Blood Pressure targets. Instruction Manual...
  • Page 2 WatchBP Office Target is designed with built-in blood pressure targets to provide convenient, accurate and reliable office blood pressure measurements. It follows the recommended guidelines of the European Society of Hypertension (ESH) , American Heart Association (AHA) , American Diabetes Association (ADA) and World Health Organization (WHO) O’Brien E, Asmar R, Beilin L, Imai Y, et al.
  • Page 3: Table Of Contents

    Table of Contents Product description Name of parts and display ......... 3–4 Viewing and deleting of measurements Components and accessories......... 5 Viewing the stored values ........12 Pulse pressure (PP) ..........12 Deleting the stored values ........13 Before using WatchBP Office Target Replacing slot-in instruction card ......
  • Page 4: Product Description

    Product description Name of parts Main Adaptor Display SYS. mmHg DIA. mmHg Instruction Cuff Socket PUL. /min M Button (Memory) Start/Stop Button Battery Compartment Mode Switch...
  • Page 5 Display Target 2 < 130/80 Target 1 < 140/90 Systolic Value Difference to Target setting Heart Beat Diastolic Value Interval Time Battery Symbol Pulse Pressure (PP) Memory Triple Mode Pulse Rate...
  • Page 6: Components And Accessories

    WatchBP Office Target and components AC Adaptor X 1 WatchBP Office Target Input: 100-240V~50/60Hz 0.7A Output: +6V 0.6A Blood Pressure Device X 1 M size (22cm~32cm) X 1 Instruction Manual X 1 L size (32cm~42cm) X 1 Quick Start Guide X 1...
  • Page 7: Before Using Watchbp Office Target

    Before using WatchBP Office Target Two ways of activating the Device Selecting the correct cuff 1) Pull out the protective strip projecting from the Two different cuff sizes are provided with the battery compartment. WatchBP Office Target: Medium and Large. Use the cuff marker to select the cuff size that best matches 2) Plug the adapter cable into the main adapter the circumference of the patient’s upper arm.
  • Page 8 Before using WatchBP Office Target (cont.) Connect the cuff to the device Fitting the cuff properly Connect the cuff to the device by inserting the cuff 1) Place the cuff over the left (right) upper arm so connector into the cuff connector socket. that the air tube and artery mark arrow point towards the lower arm.
  • Page 9 3) Wrap and tighten the cuff around the arm. 5) It may result in incorrect blood pressure readings if the cuff does not fit properly. Use a different 4) Leave a little free space between the arm of the size cuff if the range index at the end of the cuff patient and the cuff.
  • Page 10 Selecting the operation mode: Standard or Triple mode Select an operation mode Automated triple measurements There are two measuring modes to be selected, In «3» Mode, the WatchBP Office Target Standard mode (standard single measurement) or device automatically takes three consecutive Triple mode (automatic triple measurements).
  • Page 11: Selection Of Operation Modes «Single» Or «Triple» Mode («Double» Mode As Option)

    3 measurements are averaged Press «140/90» or «130/80» button SYS. On the bottom of the right hand section of the Press the button ON/OFF «140/90» or ON/OFF mmHg display shows a number to indicate which of «130/80», to select the preferred target value. The DIA.
  • Page 12: Green Light

    Traffic Light Indication in the display Red light Green light If either the measured systolic or diastolic result is If both the measured systolic and diastolic blood above the set target value (140/90 or 130/80), pressure values are below the set target values the screen will light in red with the difference to (140/90 or 130/80), the screen will light in green the desired value shown in display as number, e.g.
  • Page 13: Viewing The Stored Values

    PP (Pulse Pressure) Viewing the stored values The WatchBP Office Target device provides the pulse The device stores only the blood pressure values pressure (PP) of the patient: pulse pressure = systolic of the last measurement procedure. To view SYS. - diastolic pressure.
  • Page 14: Deleting The Stored Values

    Clearing all values Replacing the inserted Instruction Card If you are sure that you want to permanently You can replace the inserted instruction card by remove all stored values, hold down the M-button pulling it out to the side and replacing the paper. (the instrument must have been switched off Extra cards are supplied with the instrument for this SYS.
  • Page 15: Battery And Power Adaptor

    Battery and power adaptor Battery Indicator and Battery change Flat battery – replacement Low battery When the batteries are flat, the battery symbol will When the batteries are approximately ¾ empty the flash as soon as the instrument is switched on (flat battery symbol will flash as soon as the instrument battery displayed).
  • Page 16 Using a mains adaptor Please use only the Microlife adaptor (DC 6V, 600mA) supplied with the WatchBP Office Target to operate the device. 1) Plug the adapter cable into the Power Plug of the WatchBP Office Target device. 2) Plug the adaptor plug into the wall socket. When the power adaptor is connected, no battery power is consumed.
  • Page 17: Error Messages

    • by the cuff, caused for Please contact your local instance by movement or Microlife service center if the error persists. muscle tension. Repeat • the measurement, If you think the results keeping your arm still.
  • Page 18 Error Description Potential cause and Error Description Potential cause and remedy remedy «Err 6 Triple Mode There were too many «Err 3» No pressure An adequate pressure errors during the in the cuff cannot be generated measurement in in the cuff. A leak Triple mode, making it may have occurred.
  • Page 19 Error Description Potential cause and remedy «HI» Pulse or cuff The pressure in the pressure too cuff is too high (over high 300 mmHg) OR the pulse is too high (over 200 beats per minute). Relax for 5 minutes and repeat the measurement.
  • Page 20: Safety, Care, Accuracy Test And Disposal

    Safety, care, accuracy test and disposal Safety and protection Observe the storage and operating conditions This device may be used only for the purpose described in the “Technical specifications” section described in this booklet. The device comprises of this manual. of sensitive components and must be treated with caution.
  • Page 21: Accuracy Test

    Disposal An accuracy test is recommended every 2 years Batteries and electronic instruments must be or after mechanical impact (e.g. being dropped). disposed of in accordance with the locally applicable Please contact Microlife to arrange for an accuracy regulations. test.
  • Page 22: Technical Specifications

    • Pulse accuracy: ±5 % of the readout Serial number Manufacturer value • Voltage source: 4 X 1.5V Batteries; size AA • Mains adapter DC 6V, 1 A Microlife reserves the right to alter technical specfications without prior written notice.
  • Page 23: Guarantee Card

    Guarantee card This device is covered by a two-year Name: guarantee and accessories are cover by Address: a one-year guarantee from the date of purchase. This guarantee is valid only Date: on presentation of the guarantee card completed by the owner confirming date of Telephone: purchase or purchase receipt.
  • Page 24 Europe / Middle-East / Africa Asia North / Central / South America Microlife AG Microlife Corporation Microlife Medical Home Solutions, Inc. Espenstrasse 139 9F, 431, RuiGang Road, NeiHu 2801 Youngfield St., Suite 241 9443 Widnau, Switzerland Taipei, 114, Taiwan, R.O.C.

This manual is also suitable for:

Watchbp office target

Table of Contents