Taking Measurement In Auto Mode; Viewing Stored Values; Manual Mode Settings - Microlife WatchBP Office Instruction Manual

Professional automated office blood pressure monitor
Hide thumbs Also See for WatchBP Office:
Table of Contents

Advertisement

Taking measurement in AUTO Mode

Select AUTO Mode. Press the
settings of AUTO mode. The device shows all the settings and then starts counting down the
Resting Time before the first measurement. The average measurement reading is displayed
and saved after the measurements are complete.
*
If CBP measurement is enabled, the cuff pressure is hold at around 50-150mmHg for around 10 seconds to collects sufficient
pulse waves.
*
Press Start/Stop during countdown skip the counting down.
*
Pressing Start/Stop to deflation terminate the measurement while cuff is inflating or deflating. Display the results (average) if
available.

Viewing stored values

The device stores blood pressure values of the last measurement procedure in AUTO Mode.
Press the
Button to reveal the average of the measurements of AUTO mode. Continue
pressing the
Button to review individual measurements.

MANUAL Mode Settings

The measurement program in MANUAL Mode can be set to preferences. The program
includes seting the Highest Cuff Pressure and Hide Cuff Pressure during deflation.
Set Highest Inflation Pressure – press
when the device is in MANUAL mode to enter setting
of Highest Cuff Pressure. Use
to scroll among 160, 180, 200, 220 240 mmHg and
auto '- - -'. Use the
to HIDE Pressure setting.
*
When auto ' - - - ' is selected, the device automatically inflates the cuff
to a proper cuff pressure;
*
The Highest Inflation Pressure is considered as a safeguard pressure.
The device automatically inflates the cuff to the optimal cuff pressure
but not higher than the selected Highest Cuff Pressure.
Hide Cuff Pressure during deflation – This option
helps you to determinetake Korotkoff K1 and K5
sound readings without digit preference. To use the
HIDE function in MANUAL mode, push
select the setting of HIDE function and confirm the
selection by
button and finish the setting of
MANUAL mode.
10
Button to perform automatic measurements base on the
button
button or
button to confirm and move
button to
button