Observe the storage and operating conditions as described in the "Technical
speciications" section of this manual.
Protect the device from water and
moisture
Protect the device from direct
sunlight
Protect the device from extreme
heat and cold

Device cleaning and disinfecting

Use a soft cloth with one of the following recommended
cleaning solutions to wipe the exterior of the device:
• Ethyl or isopropyl alcohol (70% solution).
• Hydrogen peroxide 7.5% solution.
• Sodium hypochlorite solution (5.25-6.15%
household bleach diluted 1:500 provides >100 ppm
available chlorine)
Then wipe the exterior of the device with a soft, dry cloth.

Cuff cleaning and disinfecting

Remove the bladder. Fold and place the cuff cover
inside a washing bag. Wash the cuff cover with warm
water (43°C; 110°F) and a mild detergent in the
washing machine.
Pasteurization: wash the cuff cover in 75°C(167°F)
hot water for 30 minutes.

Accuracy test

We recommend the device to be tested for accuracy
every 2 years or after mechanical impact (e.g. Being
dropped). Please contact Microlife to arrange an
accuracy test.

Disposal

Batteries and electronic instruments must be disposed of in accordance with
the locally applicable regulations, and not as domestic waste.
WatchBP Office
Avoid proximity to electromagnetic
fields, such as those produced by
mobile phones
Never open the device
Protect the device from impact and
drops
MODE
Do not iron the cuff!
17