Taking measurement in Single or Both Arm(s) Mode

button to start the measurement when in standby mode according to the
Press the
settings in Part A) when the measurement of one arm or both arms is selected. The
device shows all the settings and then starts counting down the Resting Time before the
first measurement is taken. The average measurement reading is displayed and saved
after the measurements are complete.
The device switches the display showing
the average Systolic Blood Pressure (SYS),
Diastolic Blood Pressure(DIA), Mean Arterial
Pressure (MAP), central Systolic Blood
Pressure (cSYS), central Diastolic Blood
Pressure(cDBP), and central Pulse Pressure
(cPP) automatically if the central blood
pressure measurement is enabled.
If CBP measurement is enabled, the cuff pressure is held at
*
around 60 mmHg for around 10 seconds to collect sufficient
pulse waves.
If the difference in blood pressure readings between the two
*
arms is more than 15 mmHg for Systolic for more than three
measurements, the "IAD" icon will show with the result. The
"IAD" icon and readings of the higher arm flash.
*
Press
button during countdown to skip the countdown if
needed.
Press
button to cancel the remaining measurements at any
*
time during the measurement sequence. Displays the results
(average) if available.

Viewing stored values

The device stores blood pressure values of the last measurement(s) procedure.
button to reveal the average of the measurements. Continue pressing the
Press the
button to review individual measurements.
The device switches the display for values of the individual measurement (including SYS, DIA, MAP, cSYS, cDIA and cPP
*
values) if central blood pressure measurement is enabled.
*
Press
button to go back to standby mode.
Settings of the measurement
Part B) For measurement on one arm with one leg
When assessing brachial-ankle Pulse Wave Velocity (baPWV) and Ankle Brachial Index
(ABI) measurement (one arm and one ankle measurement is selected).
Set the Height of User (in cm) – Press the
selected for measurement, to enter the Height setting. Use the
button to scroll down the Height values. Press
use
to standby mode.
*
Height range is between 120 and 210 cm.
Only for the device version with baPWV feature.
*
WatchBP Office
button when one arm and one ankle is
button to confirm and return
button to scroll up and
13