Microlife WatchBP Office Vascular Instruction Manual page 11

Professional ofice blood pressure and cardiovascular screening monitor
Hide thumbs Also See for WatchBP Office Vascular:
Table of Contents

Advertisement

3 Set the Interval Time – between measurements - Press the
button to scroll up and use
30, 60, 120, 180, 240, 300 seconds of Interval Time. Prese
button to confirm and enter AFIB detector setting.
Set the interval time will be skipped if Number of Measurements is 1.
*
While the number of measurements is set by 6 and CBP option is on, the max
*
Interval Time is 240.
4 Set the Atrial fibrillation (AFIB) detector
– Press
the
AFIB detector ON or OFF. Prese
button to confirm.
Set the AFIB detector will be skipped if Number of
*
Measurements is 1.
5 Set the Central blood pressure (CBP)
measurement – Press
switch ON or OFF CBP measurement.
button to confirm. If the device
Prese
is upgraded with CBP then it is switched on
in default.
6 Average calculation – The device features
allow you to discard 1st measurement from
averaging while the selected number of
measurement is 3, 4, 5, or 6 measurements.
button to switch ON or OFF
Press
or
and prese
of Discarding 1
feature and finish the setting. Once you go
through the settings, the device returns to
standby.
The last settings programmed to the device are the default
*
of AUTO mode until you set the program again.
Average calculation will be skipped if Number of
*
Measurements is 1 or 2 .
button to scroll down among 15,
or button to switch the
button to
or
button to confirm the setting
st
measurement (d-1)
WatchBP Office
11
EN