Microlife WatchBP Office Central Instruction Manual

Cardiovascular screening solution for central blood pressure measurement, ankle brachial index assessment, atrial fibrillation detection, simultaneous assessment of the inter-arm, blood pressure difference
Hide thumbs Also See for WatchBP Office Central:

Advertisement

Quick Links

WatchBP Office Central
WatchBP Office Central
Cardiovascular Screening Solution for:
-Central blood pressure measurement
-Ankle Brachial Index assessment
-Atrial fibrillation detection
-Simultaneous assessment of the Inter-Arm
Blood Pressure Difference
Instruction Manual
3
EN

Advertisement

Table of Contents
loading

Summary of Contents for Microlife WatchBP Office Central

  • Page 1 WatchBP Office Central WatchBP Office Central Cardiovascular Screening Solution for: -Central blood pressure measurement -Ankle Brachial Index assessment -Atrial fibrillation detection -Simultaneous assessment of the Inter-Arm Blood Pressure Difference Instruction Manual...
  • Page 2 1. Cheng, H.M., et al., Measurement Accuracy of a Stand-Alone Oscillometric Central Blood Pressure Monitor: A Validation Report for Microlife WatchBP Office Central. Am J Hypertens. 2013. 26: 42-50.
  • Page 3: Table Of Contents

    Components and accessories ......... 8 Central pulse pressure .......... 23 Mean arterial pressure and pulse pressure ..24 Before using WatchBP Office Central Appearance of the atrial fibrillation indicator for early Selecting and fitting the correct cuff ...... 9 detection ............. 25 Selective Cuffs for upper arm and ankle ....
  • Page 4: Product Description

    Product description Name of parts I/O switch Cuff Socket (R) Main Adaptor Socket DC 7.5V 2.0A DC 7.5V 2.0A Battery Compartment Cuff Socket (L) USB Port ROUTINE AUSCULT. SCREEN Recharge Indicator Display M Button (Memory) Mode Switch Start/Stop Button ABI CENTRAL SCREEN...
  • Page 5 WatchBP Office Central Display Mean Arterial Pressure (MAP) Pulse Pressure (PP) Value for Left Arm Mode Indication Systolic Value External Power Diastolic Value Atrial Fibrillation Detection Central Systolic/ Pulse Pressure Inter-arm Difference L/R Cuff Value for Right Arm Supine Position Reminder...
  • Page 6: Components And Accessories

    Components and accessories AC Adaptor UES18LCP-075200SPA X1 WatchBP Office Central (Input: 100-240V~50/60Hz 0.5A Blood Pressure Device Output: +7.5V 2A) USB Cable Quick start guide Cuff for upper arm Product description I/O switch Cuff Socket (R) Main Adaptor Socket DC 7.5V 2.0A...
  • Page 7: Before Using Watchbp Office Central

    Before using WatchBP Office Central WatchBP Office Central Selecting the correct cuff Fitting the cuff properly Two upper arm cuffs and an ankle cuff are provided 1) Place the cuff over the left (or right) upper within the standard package of the WatchBP Office arm so that the air tube and artery mark arrow Central.
  • Page 8: Selective Cuffs For Upper Arm And Ankle

    22 - 32 cm (8.7 - 12.6 inches) 32 - 42 cm (12.6 - 16.5 inches) With air tube 200 cm With air tube 200 cm * Please contact Microlife or its distributor to purchase L-XL size or other size cuffs.
  • Page 9: Application Of The Ankle Cuff

    WatchBP Office Central 3) Wrap and tighten the cuff around the leg. Application of the ankle cuff (in ABI mode) 4) Leave a little free space between the leg of the patient and the cuff. Two fingers should fit 1) The patient has to lie down in supine position.
  • Page 10: Three Operation Modes

    Select an operation mode «ABI» Mode AUSCULT. SCREEN The WatchBP Office Central device includes three Select «ABI» Mode for Ankle-Brachial operation modes: «ABI», «CENTRAL», and «SCREEN» measurement. Select the arm with the higher (Inter-arm difference). Use the Mode Switch to select blood pressure value as determined with the the desired mode.
  • Page 11: Central» Mode

    * The user can manually set measurement intervals of Automated double measurements 15, 30, 45 or 60 seconds in CENTRAL Mode. (Please In «CENTRAL» Mode, the WatchBP Office Central refer to special functions section page 19 “Setting device automatically takes two consecutive...
  • Page 12: Screen» Mode

    Three operation modes (cont.) Simultaneous dual-arm measurements «SCREEN» Mode In «SCREEN» Mode, the WatchBP Office Central device Select «SCREEN» Mode to complete fully- measures the patient’s blood pressure on both automated triple measurements on both arms arms simultaneously, for determining the arm with...
  • Page 13: Inter-Arm Difference (Iad)

    WatchBP Office Central Automated triple measurements Determine the proper arm and inter-arm difference In «SCREEN» Mode, the WatchBP Office Central device automatically takes three consecutive If the difference of blood pressure readings between two arms is more than 20mmHg for Systolic or measurements on both arms at fixed one minute* intervals.
  • Page 14: Taking Measurements Using Watchbp Office

    Taking measurements using WatchBP Office Central «SCREEN» Mode 3) Switch to «SCREEN» Mode – Slide the Mode Switch to «SCREEN» Mode. 1) Turn on the power – Turn on the device by switching the I/O switch at the back of the device to the ON position.
  • Page 15: Taking Less Than Three Measurements

    WatchBP Office Central Taking less than three measurements 6) Three consecutive measurements – The device will take three consecutive measurements with The measurement sequence can be stopped at one minute intervals between measurements. anytime by pressing the Start/Stop Button in «SCREEN»...
  • Page 16: Central» Mode

    Taking measurements using WatchBP Office Central (cont.) «CENTRAL» Mode 3) Switch to «CENTRAL» Mode – Slide the Mode Switch to «CENTRAL» Mode. 1) Turn on the power – Turn on the device by switching the I/O switch at the back of the device to the ON position.
  • Page 17: Central" Mode

    WatchBP Office Central 5) Press the M Button to switch between left(L) and 7) Automatic two measurements – The device right(R). Press the Start/Stop Button to save the will take two consecutive measurements with setting. 15 second intervals in between on default.
  • Page 18: Skipping The Countdown Time

    Taking measurements using WatchBP Office Central (cont.) Skipping the countdown time Setting measuring intervals in «CENTRAL» Mode The 60 seconds countdown before measurement in «SCREEN» Mode and 15 seconds in «CENTRAL» Mode 1) Pressing and holding the Start/Stop Button for 3 can be skipped by pressing the Start/Stop Button.
  • Page 19: Abi» Mode

    WatchBP Office Central «ABI» Mode 1) Turn on the power – Turn on the device by 4) Apply the cuff* – Properly fit the arm cuff to the switching the I/O switch at the back of the upper arm and the ankle cuff to the leg of the device into the ON position.
  • Page 20 Taking measurements using WatchBP Office Central (cont.) 6) Start measurements – Press the Start/Stop 8) Display of the measurement values– Button to start the measurement. The measurement values will be displayed and the Ankle-brachial index is automatically calculated when the measurement is completed.
  • Page 21: Special Functions

    Am J Hypertens, 2012. 25: 542-8. 2. Cheng, H.M., et al., Measurement accuracy of a stand-alone oscillometric central blood pressure monitor: a validation report for Microlife WatchBP Office Central. Am J Hypertens, 2013. 26: 42-50.
  • Page 22: Mean Arterial Pressure And Pulse Pressure

    PP (Pulse Pressure) MAP (Mean Arterial Pressure) This measurement device provides the pulse The WatchBP Office Central device measures the true pressure (PP) of the patient: pulse pressure = systolic mean arterial pressure (MAP). Each measurement includes a single MAP value. The average - diastolic pressure.
  • Page 23: Appearance Of The Atrial Fibrillation Indicator For Early Detection

    WatchBP Office Central Appearance of the atrial fibrillation indicator for early detection The WatchBP Office Central is designed to screen for asymptomatic atrial fibrillation during blood pressure SCREEN measurements in the «SCREEN» Mode. If two out of three measurements in the «SCREEN» Mode detect atrial fibrillation, the Afib icon is displayed.
  • Page 24: About Atrial Fibrillation

    Special Function (cont.) About Atrial Fibrillation (Afib) Afib is a common heart rhythm problem. It affects A doctor can confirm the presence of Afib by using a more than 2 million people in North America. It is 12-lead ECG. Sometimes Afib is present incidentally. more common in older age and it is found in 18% Therefore, a doctor may not see it on regularly individuals aged 85 years and older.
  • Page 25: Abi (Ankle Brachial Index)

    WatchBP Office Central ABI (Ankle Brachial Index) A low ankle-brachial index (ABI) is indicative of systemic vascular disease, and places a patient in the The systolic blood pressure values of a person’s arm high-risk category. and leg need to be measured In order to determine ABI (Ankle Brachial Index).
  • Page 26: Hiding Measurement Values

    1) Activate the «Hide» Function – Press M button; L or R flashes; keep pressing and holding the M The WatchBP Office Central features a Hide function Button for 7 more seconds until On or Off flashes. in order to prevent unnecessary elevated blood...
  • Page 27: Viewing And Transferring Measurements

    Viewing and transferring measurement readings WatchBP Office Central «SCREEN» Mode 1) Viewing the average of all measurements When the last measurement is performed in seconds... «SCREEN» mode, press the M button displays the average of the 3 measurements of the «SCREEN»...
  • Page 28: Central» Mode

    Viewing and transferring Measurements (cont.) «CENTRAL» Mode 3) Viewing individual measurements – All individual readings can be viewed by 1) Viewing the average of all measurements repeatedly pressing the M Button. The display When the last measurement is performed in will flash a number before the reading «CENTRAL»...
  • Page 29: Abi» Mode

    WatchBP Office Central «ABI» Mode 3) Viewing MAP and PP values – Ten seconds after display of the average blood 1) When the last measurement is performed in pressure value, the values for Pulse Pressure (PP) «ABI» mode, press the M button displays the and Mean Arterial Pressure (MAP) are displayed.
  • Page 30: Transferring Measurements

    Viewing and transferring Measurements (cont.) Transferring measurements Transferring data to the computer Installation of the software program 1) Start the software program and connect the device to the computer using the cable 1) Put the CD in the CD-ROM drive of your supplied.
  • Page 31: Appendix

    Rechargeable Battery Using a power adaptor The device is equipped with a built-in, rechargeable Please only use the Microlife adaptor supplied with Ni-MH battery pack which delivers up to 400~500 the device to recharge the device. measurement cycles. The batteries can be recharged 1) Plug the adapter cable into the Power Plug of the while the power adaptor is plugged in.
  • Page 32: Troubleshooting

    Leakage of the tube or Check for cracks on the tube or the bladder. Please bladder contact Microlife customer service for this issue. No result displayed The Hide function is Disable “hide” function between measurements...
  • Page 33: Error Messages

    • instance by movement or Please contact your local muscle tension. Repeat Microlife service center if the error persists. the measurement, • keeping your arm still. If you think the results are unusual, please read...
  • Page 34 Error messages Err 11 Signal too The pulse signals on Err 3 No pressure in An adequate pressure weak during the cuff are too weak. the cuff cannot be generated central blood Re-position the cuff and in the cuff. A leak pressure repeat the measurement.
  • Page 35 WatchBP Office Central Err 13 Cuff pressure An adequate pressure Pulse or cuff The pressure in the cuff errors during cannot be generated in pressure too is too high (over 300 Central blood the cuff. A leak may have high...
  • Page 36: Safety, Care, Accuracy Test And Disposal

    Safety, care, accuracy test and disposal Read the instruction manual carefully before using this device, especially the safety instructions, and keep the instruction manual for future use. Do not use the instrument close to strong electromagnetic fields such as mobile telephones or radio installations Observe the storage and operating conditions Safety and protection described in the “Technical specifications”...
  • Page 37 Do not wash and scrub cuff Do not iron the cuff! Accuracy test We recommend the WatchBP Office Central device be tested for accuracy every 2 years or after mechanical impact (e.g. being dropped). Please contact Microlife to arrange for an accuracy test.
  • Page 38: Technical Specifications

    • Pulse accuracy: ±5 % of the readout For indoor use value • Rechargeable battery pack; Voltage source: 4.8V C3500 mAh Microlife reserves the right to alter technical specifications without prior written notice. • Mains adapter DC 7.5V, 2 A...
  • Page 39 This guarantee is valid only Date: on presentation of the guarantee card completed by the owner confirming date of Telephone: purchase or purchase receipt. E-mail: Product model: WatchBP Office Central REF: TWIN200 CBP...
  • Page 41 Tel. +1 303 274 2277 Tel +1 727 442 5353 Fax +41 71 727 7011 Email: Email: watchbp@microlife.ch watchbp@microlife.com.tw Fax +1 303 274 2244 Fax +1 727 442 5377 www.watchbp.com www.watchbp.com Email: watchbp@mimhs.com Email: msa@microlifeusa.com www.watchbp.com www.watchbp.com IB WatchBP Office Central EN 4716...