Agilent Technologies OmniBER OTN Series Installation Manual page 30

Communications performance analyzer
Hide thumbs Also See for OmniBER OTN Series:
Table of Contents

Advertisement

Installation.book Page 27 Monday, January 12, 2004 1:13 PM
Installation Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
To clean the optical connectors
1 Disconnect the instrument from the Power Line or switch off
the laser transmitter before commencing this cleaning
procedure.
2 Remove the adapters from the optical IN and OUT ports by
flipping back the lever on the optical adapter.
3 Using the blow brush with the brush removed blow through
the ferrule of the standard flexible connector and the
adapter.
4 If the optical fiber of the fixed connector requires further
cleaning this entails disassembly of the module. This should
be carried out only by suitably trained service personnel.
5 Apply some isopropyl alcohol to a piece of the cleaning paper
and clean the barrel of the adapter. Using a new piece of
cleaning paper, clean the face of the adapter. Repeat this
operation, using a new piece of cleaning paper each time.
6 Use a blow brush or compressed air to remove any particles
of cleaning paper which may be present.
7 These procedures also apply to any Optical patch cords or
attenuators used to avoid subsequent contamination/damage
to the Optical ports.
Replace the adapters in the optical connector. Secure in place
by clicking the retaining lever back into position.
1
Introduction
27